Adult Invasive Mechanical Ventilation
A mechanics-first clinical guide to adult invasive ventilation, from pressure interpretation and mode structure to patient–ventilator interaction and troubleshooting.
On this page
- Scope and terminology
- Respiratory pressures and mechanics
- Structure of a ventilator mode
- Continuous mandatory ventilation
- Intermittent mandatory ventilation
- Spontaneous ventilatory support
- Adaptive and hybrid modes
- Specialized and device-dependent modes
- Manufacturer crosswalk
- Physiology-guided clinical application
- Waveforms and patient–ventilator interaction
- Troubleshooting
Scope and terminology
This guide addresses positive-pressure invasive ventilation through an artificial airway in adults receiving intensive care. Paediatric and neonatal ventilation, non-invasive and home ventilation, and anaesthesia-workstation operation are outside its scope.
Mode labels describe control logic, not an indication or a complete prescription. Select targets and alarms for the patient, disease, monitoring capability, and device; this guide does not provide universal starting settings or claim that one control variable is superior.
Respiratory pressures and mechanics
With respiratory-system inertance neglected, the linear single-compartment equation of motion during inspiration is Pvent(t) + Pmus(t) = PEEPtotal + V(t)/Crs + Rinsp × V̇insp(t). In this applied-pressure convention, inspiratory Pmus is positive: it is the pressure contribution applied by the inspiratory muscles, not the signed pleural-pressure change; Pmus = 0 during passive ventilation.
Pressure quantities Pvent, Pmus, Paw, Palv, Ppl, PEEPset, PEEPtotal, Ppeak, Pplat, ΔP, Pmean, and PL are expressed in cm H₂O. V(t) is the volume change above the end-expiratory reference at PEEPtotal, where V(0) = 0; VT is the end-inspiratory volume change, and both use L or mL consistently. V̇insp is in L/s; Crs is in L/cm H₂O or mL/cm H₂O; Rinsp and Rexp are in cm H₂O·s/L; τinsp and τexp are in seconds.
Static partitioning requires a passive patient, no relevant leak, stable volume, and zero flow during a valid end-inspiratory occlusion or end-expiratory occlusion. Known constant inspiratory flow is required for the simple Rinsp equation. Coughing, active inspiration or expiration, leaks, and unstable flow invalidate or distort these estimates.
During assisted ventilation, patient effort and ventilator pressure act together. Inspiratory muscle pressure can lower measured airway pressure while increasing flow, volume, or transpulmonary distending pressure; continued effort during an occlusion can make plateau pressure or total PEEP misleading.
| Pressure | Definition and relation | Valid measurement and limits |
|---|---|---|
| Peak inspiratory pressure (Ppeak) | The maximum airway pressure during inspiration. With passive constant-flow inflation, it contains set and intrinsic PEEP plus elastic and resistive pressure. | Interpret with the simultaneous flow waveform and Pplat. A change in flow or patient effort changes Ppeak even if respiratory-system mechanics are unchanged. |
| Plateau pressure (Pplat) | Airway pressure after pressure equilibrates during zero flow at end inspiration; it approximates respiratory-system elastic pressure plus total PEEP, not lung stress alone. | Use a valid end-inspiratory occlusion during passive inflation with no leak. Confirm a stable plateau and exclude inspiratory or expiratory muscle activity. |
| Resistive pressure | The pressure required to generate flow through the tube, circuit, airways, and flow-dependent tissue resistance; under constant flow it is approximated by Ppeak − Pplat. | Compare breaths at the same flow and volume. The difference cannot be interpreted as resistance when flow is changing or effort distorts airway pressure. |
| Elastic pressure | The pressure above total PEEP required to deliver VT against combined lung and chest-wall elastance; under passive static conditions it is Pplat − PEEPtotal. | It is a respiratory-system quantity. Chest-wall stiffness and abdominal pressure can raise it without an equivalent rise in lung transpulmonary pressure. |
| Set, total, and intrinsic PEEP | Set PEEP is the ventilator baseline; total PEEP is static end-expiratory alveolar pressure; intrinsic PEEP (auto-PEEP) is PEEPtotal − PEEPset. | Measure total PEEP with a valid end-expiratory occlusion in a passive patient without leak. Expiratory flow that has not returned to zero suggests incomplete emptying but does not quantify regional gas trapping. |
| Driving pressure (ΔP) | For passive tidal inflation, ΔP = Pplat − PEEPtotal = VT/Crs. It represents the tidal elastic pressure of the whole respiratory system. | Use Pplat and total, not merely set, PEEP from valid static measurements. Comparisons are unreliable when VT, effort, chest-wall mechanics, or occlusion quality changes. |
| Mean airway pressure (Pmean) | The time-average of airway pressure over the complete respiratory cycle: Pmean = (1/T)∫₀ᵀ Paw(t)dt. Here T is the duration of the complete respiratory cycle in seconds. | It depends on the pressure waveform, PEEP, inspiratory duration, frequency, flow pattern, and spontaneous effort. It is not interchangeable with Pplat or ΔP. |
| Transpulmonary pressure (PL) | PL = Palv − Ppl: alveolar minus pleural pressure. It separates lung distension from chest-wall load; airway Pplat alone cannot do this. | Pleural pressure is commonly estimated with calibrated oesophageal pressure, with position and regional limitations. Specify whether an absolute static value or a tidal swing is used; no universal PL target is inferred here. |
Peak inspiratory pressure (Ppeak)
- Definition and relation
- The maximum airway pressure during inspiration. With passive constant-flow inflation, it contains set and intrinsic PEEP plus elastic and resistive pressure.
- Valid measurement and limits
- Interpret with the simultaneous flow waveform and Pplat. A change in flow or patient effort changes Ppeak even if respiratory-system mechanics are unchanged.
Plateau pressure (Pplat)
- Definition and relation
- Airway pressure after pressure equilibrates during zero flow at end inspiration; it approximates respiratory-system elastic pressure plus total PEEP, not lung stress alone.
- Valid measurement and limits
- Use a valid end-inspiratory occlusion during passive inflation with no leak. Confirm a stable plateau and exclude inspiratory or expiratory muscle activity.
Resistive pressure
- Definition and relation
- The pressure required to generate flow through the tube, circuit, airways, and flow-dependent tissue resistance; under constant flow it is approximated by Ppeak − Pplat.
- Valid measurement and limits
- Compare breaths at the same flow and volume. The difference cannot be interpreted as resistance when flow is changing or effort distorts airway pressure.
Elastic pressure
- Definition and relation
- The pressure above total PEEP required to deliver VT against combined lung and chest-wall elastance; under passive static conditions it is Pplat − PEEPtotal.
- Valid measurement and limits
- It is a respiratory-system quantity. Chest-wall stiffness and abdominal pressure can raise it without an equivalent rise in lung transpulmonary pressure.
Set, total, and intrinsic PEEP
- Definition and relation
- Set PEEP is the ventilator baseline; total PEEP is static end-expiratory alveolar pressure; intrinsic PEEP (auto-PEEP) is PEEPtotal − PEEPset.
- Valid measurement and limits
- Measure total PEEP with a valid end-expiratory occlusion in a passive patient without leak. Expiratory flow that has not returned to zero suggests incomplete emptying but does not quantify regional gas trapping.
Driving pressure (ΔP)
- Definition and relation
- For passive tidal inflation, ΔP = Pplat − PEEPtotal = VT/Crs. It represents the tidal elastic pressure of the whole respiratory system.
- Valid measurement and limits
- Use Pplat and total, not merely set, PEEP from valid static measurements. Comparisons are unreliable when VT, effort, chest-wall mechanics, or occlusion quality changes.
Mean airway pressure (Pmean)
- Definition and relation
- The time-average of airway pressure over the complete respiratory cycle: Pmean = (1/T)∫₀ᵀ Paw(t)dt. Here T is the duration of the complete respiratory cycle in seconds.
- Valid measurement and limits
- It depends on the pressure waveform, PEEP, inspiratory duration, frequency, flow pattern, and spontaneous effort. It is not interchangeable with Pplat or ΔP.
Transpulmonary pressure (PL)
- Definition and relation
- PL = Palv − Ppl: alveolar minus pleural pressure. It separates lung distension from chest-wall load; airway Pplat alone cannot do this.
- Valid measurement and limits
- Pleural pressure is commonly estimated with calibrated oesophageal pressure, with position and regional limitations. Specify whether an absolute static value or a tidal swing is used; no universal PL target is inferred here.
| Mechanical quantity | Equation, units, and conditions | Interpretation |
|---|---|---|
| Static respiratory-system compliance | Crs = VT / (Pplat − PEEPtotal), expressed in mL/cm H₂O or L/cm H₂O, using passive zero-flow measurements without leak. | A change can arise from lung volume, recruitment or overdistension, oedema, pleural or abdominal factors, chest-wall mechanics, position, or invalid measurement. |
| Inspiratory respiratory-system resistance (Rinsp) | Rinsp = (Ppeak − Pplat) / V̇insp, expressed in cm H₂O·s/L, using passive volume-controlled inflation with known constant inspiratory flow. | A rise prompts assessment of the artificial airway, secretions, bronchospasm, circuit resistance, and flow. Compare only under matched flow conditions. |
| Inspiratory and expiratory time constants | τinsp = Rinsp × Crs and τexp = Rexp × Crs, both expressed in seconds. In a passive linear single-compartment model, one τ corresponds to about 63% of the applicable volume change after a pressure step. | Rexp is not assumed to equal Rinsp. τexp describes passive expiration with Pmus = 0 and no leak; it reflects expiratory airway and ventilator expiratory-circuit resistance, whereas τinsp uses the separately measured Rinsp. Heterogeneous lungs contain multiple time constants, so either global value is an approximation. |
Static respiratory-system compliance
- Equation, units, and conditions
- Crs = VT / (Pplat − PEEPtotal), expressed in mL/cm H₂O or L/cm H₂O, using passive zero-flow measurements without leak.
- Interpretation
- A change can arise from lung volume, recruitment or overdistension, oedema, pleural or abdominal factors, chest-wall mechanics, position, or invalid measurement.
Inspiratory respiratory-system resistance (Rinsp)
- Equation, units, and conditions
- Rinsp = (Ppeak − Pplat) / V̇insp, expressed in cm H₂O·s/L, using passive volume-controlled inflation with known constant inspiratory flow.
- Interpretation
- A rise prompts assessment of the artificial airway, secretions, bronchospasm, circuit resistance, and flow. Compare only under matched flow conditions.
Inspiratory and expiratory time constants
- Equation, units, and conditions
- τinsp = Rinsp × Crs and τexp = Rexp × Crs, both expressed in seconds. In a passive linear single-compartment model, one τ corresponds to about 63% of the applicable volume change after a pressure step.
- Interpretation
- Rexp is not assumed to equal Rinsp. τexp describes passive expiration with Pmus = 0 and no leak; it reflects expiratory airway and ventilator expiratory-circuit resistance, whereas τinsp uses the separately measured Rinsp. Heterogeneous lungs contain multiple time constants, so either global value is an approximation.
Structure of a ventilator mode
Describe a mode by its control variable, breath sequence, and targeting scheme, then define how each breath starts, is limited, and ends. A commercial label is not enough to establish these properties.
| Mode element | Question | Clinical meaning |
|---|---|---|
| Control variable | Is pressure or volume/flow constrained during inspiration? | The unconstrained variable changes with resistance, compliance, effort, leak, and the chosen waveform. |
| Breath sequence | Are breaths mandatory, spontaneous, or both? | CMV contains mandatory breaths; IMV combines mandatory and spontaneous breaths; CSV contains spontaneous breaths. Patient triggering does not by itself make a breath spontaneous. |
| Targeting scheme | Which targets are fixed, adapted between breaths, proportional to a signal, or governed by a higher controller? | The targeting scheme explains which variables the algorithm may change and which measured signal drives that change. |
| Trigger | What starts inspiration? | A time signal or a validated patient signal such as pressure, flow, or diaphragm electrical activity can initiate a breath. |
| Limit | Which variable is constrained during inspiration without ending it? | Pressure, flow, or volume limits shape delivery and may activate protective boundaries; a limit is not necessarily the cycling criterion. |
| Cycle | What ends inspiration? | Time, flow, volume, pressure, or an algorithm can terminate inspiration and thereby change synchrony and expiratory time. |
Control variable
- Question
- Is pressure or volume/flow constrained during inspiration?
- Clinical meaning
- The unconstrained variable changes with resistance, compliance, effort, leak, and the chosen waveform.
Breath sequence
- Question
- Are breaths mandatory, spontaneous, or both?
- Clinical meaning
- CMV contains mandatory breaths; IMV combines mandatory and spontaneous breaths; CSV contains spontaneous breaths. Patient triggering does not by itself make a breath spontaneous.
Targeting scheme
- Question
- Which targets are fixed, adapted between breaths, proportional to a signal, or governed by a higher controller?
- Clinical meaning
- The targeting scheme explains which variables the algorithm may change and which measured signal drives that change.
Trigger
- Question
- What starts inspiration?
- Clinical meaning
- A time signal or a validated patient signal such as pressure, flow, or diaphragm electrical activity can initiate a breath.
Limit
- Question
- Which variable is constrained during inspiration without ending it?
- Clinical meaning
- Pressure, flow, or volume limits shape delivery and may activate protective boundaries; a limit is not necessarily the cycling criterion.
Cycle
- Question
- What ends inspiration?
- Clinical meaning
- Time, flow, volume, pressure, or an algorithm can terminate inspiration and thereby change synchrony and expiratory time.
Continuous mandatory ventilation
VC-CMV
Every breath is mandatory; the selected volume and flow pattern are the primary constrained variables, including for patient-triggered breaths.
Airway pressure changes with resistance, compliance, total PEEP, flow, and respiratory effort.
Patient inspiration can lower the pressure waveform despite substantial muscle pressure; a low displayed pressure does not prove low transpulmonary distension.
Review expired volume, Ppeak, a valid Pplat and total PEEP when measurable, flow and pressure waveforms, gas exchange, effort, and haemodynamics after changes.
PC-CMV
Every breath is mandatory; inspiratory pressure and timing are constrained, while flow is variable and commonly decelerating.
VT changes with the pressure gradient, respiratory-system mechanics, available inspiratory time, leak, and respiratory effort.
Patient inspiration can increase flow and VT without raising the set airway pressure, so the pressure trace alone can underestimate total inspiratory load.
Review expired VT and minute ventilation, flow decay, total PEEP, effort, gas exchange, and haemodynamics whenever mechanics or spontaneous activity changes.
Intermittent mandatory ventilation
SIMV combines mandatory breaths with spontaneous breaths between them. Evaluate each breath type separately; SIMV is not inherently a liberation strategy or a guarantee of appropriate work of breathing.
VC-SIMV
Mandatory breaths use volume control. Spontaneous breaths depend on the patient and any separate pressure support; assess the mandatory flow/volume pattern and the spontaneous trigger, cycling, effort, and contribution to total ventilation.
PC-SIMV
Mandatory breaths use pressure control. Track their delivered VT separately from supported or unsupported spontaneous breaths, and check whether changing effort alters total ventilation or creates over-assistance.
Spontaneous ventilatory support
In spontaneous support modes, the patient initiates inspiration, while breath duration is usually determined by inspiratory flow. Effective ventilation depends on respiratory drive, respiratory-muscle capacity, trigger settings, pressure rise time, cycling to expiration, respiratory mechanics, leak, and the availability of backup ventilation.
- Pressure support ventilation (PSV)
Each patient-triggered breath receives a set level of pressure support and usually cycles to expiration when inspiratory flow falls to a predefined threshold. Match trigger sensitivity, rise time, and cycling criteria to patient effort and respiratory mechanics. Monitor for ineffective efforts, excessive VT, premature cycling, and delayed cycling.
- Invasive CPAP
CPAP maintains continuous positive airway pressure without additional inspiratory assistance. It requires adequate spontaneous ventilation, reliable apnoea detection, and appropriately configured backup ventilation.
- Balance of support
Insufficient support increases respiratory effort and inspiratory-flow demand. Excessive support may reduce respiratory drive, increase VT, cause delayed cycling, and mask respiratory-muscle weakness. Assess patient effort, ventilation, comfort, and patient–ventilator synchrony together.
Adaptive and hybrid modes
In these modes, the ventilator automatically adjusts selected ventilation parameters using measured feedback. Interpret them according to the breath type, feedback signal, selected targets, active limits, and the algorithm’s response to changes in respiratory effort and lung mechanics.
- PRVC-type modes with tidal-volume targeting
Mandatory breaths are pressure controlled, while the ventilator adjusts inspiratory pressure between breaths to approach the set VT. The target volume may not be achieved when pressure limits are reached or when leaks, respiratory mechanics, or patient effort change.
- AutoFlow
AutoFlow adapts inspiratory pressure in modes with a set tidal volume and delivers inspiration with a decelerating flow pattern. Assess actual VT, airway pressure, flow, and patient–ventilator interaction. Implementation and available settings depend on the ventilator model.
- Volume-targeted pressure support
During spontaneous breaths, the ventilator adjusts pressure support between breaths to approach the set VT. Stronger patient effort may reduce ventilator pressure, while weaker effort may increase it; therefore, VT should be interpreted together with patient effort.
- Adaptive support ventilation: ASV
ASV automatically selects a combination of respiratory rate, VT, and pressure support based on the set minute ventilation, measured respiratory mechanics, and spontaneous patient activity. Assess the resulting breath sequence, VT, rate, pressure, patient effort, and active safety limits.
- INTELLiVENT-ASV
INTELLiVENT-ASV extends ASV with automated control of ventilation and oxygenation using the selected clinical profile and measured signals. Confirm the selected targets, signal quality, controller status, active limits, and backup behaviour.
Specialized and device-dependent modes
The availability and implementation of these modes depend on the ventilator model and its software configuration. Identify each mode by its control principle, patient signal, breath sequence, and backup logic rather than by its name alone.
- Time-cycled two-level pressure modes: APRV and BiLevel-type modes
The ventilator alternates between upper and lower pressure levels at predefined time intervals, while spontaneous breathing may be permitted at both levels. Assess the pressure levels, duration of each phase, release flow and volume, support of spontaneous breaths, total minute ventilation, gas exchange, and haemodynamics. Modes with similar names may differ in their algorithms and available settings.
- Minimum minute ventilation: MMV
MMV automatically increases the mandatory component of ventilatory support when measured minute ventilation falls below a predefined minimum and reduces it as adequate spontaneous ventilation returns. Confirm how the ventilator calculates minute ventilation, counts spontaneous breaths, and limits the additional support.
- Proportional ventilatory assistance: PAV+
PAV+ adjusts inspiratory assistance in proportion to estimated patient effort, using measured flow and volume together with estimated resistive and elastic loads of the respiratory system. Confirm that measurements are reliable, significant leaks are absent, and artificial-airway data are correct. Availability, operating limits, and backup behaviour depend on the specific ventilator.
- Neurally adjusted ventilatory assist: NAVA
NAVA uses the electrical activity of the diaphragm—EAdi—measured through a dedicated oesophageal catheter to trigger and cycle breaths and to scale assistance proportionally. Confirm catheter position and signal quality. Continue to assess airway pressure, VT, respiratory rate, gas exchange, and patient effort.
- Automatic tube compensation: ATC
ATC estimates the pressure drop across an endotracheal or tracheostomy tube using its type, internal diameter, and measured flow, and adds pressure to partially offset this resistance. It does not compensate for resistance caused by accumulated secretions, partial obstruction or kinking of the tube, bronchospasm, or the breathing circuit.
Manufacturer crosswalk
This table maps only behavior verified in the cited documents. Exact revisions are listed in the source metadata; a dash means that the current evidence set does not support a crosswalk entry, not that the function is unavailable.
| Generic function | Dräger | Hamilton | Important implementation distinction |
|---|---|---|---|
| Adaptive mandatory volume targeting | VC-CMV + AutoFlow | APVcmv / (S)CMV+ | Dräger exposes AutoFlow as an additional setting in its volume-controlled mode family; for CMV + AutoFlow, breath taxonomy classifies pressure control with adaptive targeting (PC-CMVa). Hamilton APVcmv also targets mandatory volume with adaptive pressure control, but the names are not interchangeable. |
| Adaptive spontaneous volume targeting | — | VS | VS applies a volume target to spontaneous breaths; it is distinct from adaptive mandatory ventilation. |
| Adaptive pattern and gas-exchange controllers | — | ASV / INTELLiVENT-ASV | ASV selects the ventilatory pattern from physiological input; INTELLiVENT-ASV adds clinician-bounded carbon-dioxide elimination and oxygenation controllers. |
| Time-cycled two-level pressure | PC-APRV | DuoPAP / APRV | Spontaneous-breath support, transition timing, and release behavior remain implementation-specific. |
| Minimum minute ventilation | VC-MMV | — | The cited implementation combines measured minute volume, volume-controlled mandatory breaths, and spontaneous activity; it is not a generic liberation protocol. |
| Artificial-airway resistance compensation | ATC | TRC | Both are device estimates based on tube and flow information; their settings and calculations are platform-specific. |
Adaptive mandatory volume targeting
- Dräger
- VC-CMV + AutoFlow
- Hamilton
- APVcmv / (S)CMV+
- Important implementation distinction
- Dräger exposes AutoFlow as an additional setting in its volume-controlled mode family; for CMV + AutoFlow, breath taxonomy classifies pressure control with adaptive targeting (PC-CMVa). Hamilton APVcmv also targets mandatory volume with adaptive pressure control, but the names are not interchangeable.
Adaptive spontaneous volume targeting
- Dräger
- —
- Hamilton
- VS
- Important implementation distinction
- VS applies a volume target to spontaneous breaths; it is distinct from adaptive mandatory ventilation.
Adaptive pattern and gas-exchange controllers
- Dräger
- —
- Hamilton
- ASV / INTELLiVENT-ASV
- Important implementation distinction
- ASV selects the ventilatory pattern from physiological input; INTELLiVENT-ASV adds clinician-bounded carbon-dioxide elimination and oxygenation controllers.
Time-cycled two-level pressure
- Dräger
- PC-APRV
- Hamilton
- DuoPAP / APRV
- Important implementation distinction
- Spontaneous-breath support, transition timing, and release behavior remain implementation-specific.
Minimum minute ventilation
- Dräger
- VC-MMV
- Hamilton
- —
- Important implementation distinction
- The cited implementation combines measured minute volume, volume-controlled mandatory breaths, and spontaneous activity; it is not a generic liberation protocol.
Artificial-airway resistance compensation
- Dräger
- ATC
- Hamilton
- TRC
- Important implementation distinction
- Both are device estimates based on tube and flow information; their settings and calculations are platform-specific.
Physiology-guided clinical application
Select the mode according to the main physiological problem and define in advance how the response will be assessed. A mode is appropriate only when the ventilation it actually delivers remains consistent with the patient-specific strategy as respiratory mechanics and effort change.
- ARDS and acute hypoxaemia
Use a mode that reliably delivers the individualized lung-protective strategy and allows assessment of expired VT, reliably measured Pplat and total PEEP, ΔP, gas exchange, respiratory effort, and haemodynamics. The mode name itself does not provide lung protection.
- Obstructive disease and dynamic hyperinflation
Provide sufficient expiratory time and confirm that expiratory flow returns to zero. Distinguish set, total, and intrinsic PEEP; assess expiratory flow limitation, trigger workload, and haemodynamics. Changing the mode without adjusting expiratory time or minute ventilation will not correct dynamic hyperinflation.
- Marked respiratory effort or dyssynchrony
Identify pain, anxiety, fever, acidosis, hypoxaemia, and other causes of increased respiratory drive. Assess work of breathing, respiratory mechanics, and whether support matches patient demand. Strong inspiratory effort may reduce airway pressure while increasing transpulmonary pressure. Adjust triggering, inspiratory flow, rise time, cycling to expiration, and support only after identifying the mechanism, then reassess.
- Neurological impairment
Assess gas exchange, airway protection, and the ability to initiate and sustain ventilation separately. Level of consciousness, sedation, seizures, respiratory drive, cough, secretion burden, intracranial pressure, and cerebral perfusion determine the need for reliable mandatory or backup ventilation and readiness for extubation.
- Liberation from mechanical ventilation
When readiness criteria are met, perform a structured assessment and a spontaneous breathing trial. During assisted ventilation, ensure that low support is not associated with excessive work of breathing and that high support is not suppressing spontaneous activity. Before extubation, also assess airway protection, cough, secretion burden, and the risk of extubation failure.
Waveforms and patient–ventilator interaction
Interpret pressure, flow, and volume together with the patient’s clinical condition. Confirm the likely cause before changing settings and reassess the response afterward.
| Problem | Pattern | Likely causes | What to check | Possible adjustment | After adjustment |
|---|---|---|---|---|---|
| Ineffective inspiratory effort | A pressure dip or expiratory-flow deflection does not trigger an assisted breath. | Intrinsic PEEP, insensitive trigger, weak effort, over-assistance, sedation, or leak. | Correlate patient effort with the waveforms; assess total PEEP, trigger function, leak, and respiratory drive. | Trigger sensitivity, expiratory time, external PEEP, support level, and reversible causes of weak effort. | Reliable triggering without auto-triggering, fewer ineffective efforts, and stable VT and ventilation. |
| Double triggering and breath stacking | Two breaths occur with little or no expiration between them, increasing total VT. | Prolonged neural inspiration, high effort, inadequate flow, premature cycling, or reverse triggering. | Assess stacked VT, inspiratory duration, flow demand, effort, pain, acidosis, and sedation. | Inspiratory flow, pressure rise, inspiratory time or cycling criterion, and support level. | Separated breaths, acceptable VT and pressures, lower effort, and preserved gas exchange. |
| Insufficient inspiratory flow | During volume control, the pressure curve becomes concave and inspiratory effort persists. | Insufficient flow, pressure rise, or support; pain, fever, anxiety, acidosis, or high ventilatory demand. | Compare delivered flow with patient demand and identify clinical causes of increased respiratory drive. | Flow pattern and peak flow, pressure rise, inspiratory time, support, and treatment of the cause. | Lower effort and distress, smoother pressure contour, acceptable VT and pressures, and adequate expiratory time. |
| Premature cycling | Ventilator inspiration ends before patient effort, often causing immediate retriggering. | High flow-cycle threshold, short inspiratory time, or insufficient flow or support. | Compare the end of ventilator inspiration with patient effort; check leak, cycling criterion, inspiratory time, and VT. | Cycling criterion, inspiratory time, inspiratory flow, and support level. | Better breath matching without excessive VT, double triggering, or shortened expiration. |
| Delayed cycling | Inspiratory flow continues after patient effort ends, with active expiration or a late pressure rise. | Low flow-cycle threshold, excessive support, leak, long inspiratory time, or obstruction. | Assess expiratory effort, leak, flow decay, respiratory mechanics, and support level. | Cycling criterion, inspiratory time, leak correction, pressure support, and pressure-rise setting. | Timely expiration, less expiratory effort, and adequate VT and ventilation. |
| Incomplete expiration | Expiratory flow does not return to zero before the next inspiration. | Obstruction, long time constant, high minute ventilation, short expiratory time, or active expiration. | Assess resistance, ventilatory demand, expiratory time, haemodynamics, and total PEEP. | Increase expiratory time; adjust rate, inspiratory flow or time, support, and treatment of obstruction. | Improved lung emptying, lower total PEEP, better triggering, gas exchange, and haemodynamics. |
| Changes in Ppeak and Pplat | Ppeak rises with stable Pplat, or both pressures rise together. | An isolated Ppeak rise suggests increased resistance; a Pplat rise suggests reduced compliance or increased total PEEP. | Repeat passive measurements and inspect the tube, circuit, airways, chest wall, lung volume, and patient position. | Treat airway or circuit obstruction; adjust VT, PEEP, position, and treatment of the underlying disease. | Repeat Ppeak, Pplat, total PEEP, Crs, and Rinsp and assess ventilation and haemodynamics. |
Ineffective inspiratory effort
- Pattern
- A pressure dip or expiratory-flow deflection does not trigger an assisted breath.
- Likely causes
- Intrinsic PEEP, insensitive trigger, weak effort, over-assistance, sedation, or leak.
- What to check
- Correlate patient effort with the waveforms; assess total PEEP, trigger function, leak, and respiratory drive.
- Possible adjustment
- Trigger sensitivity, expiratory time, external PEEP, support level, and reversible causes of weak effort.
- After adjustment
- Reliable triggering without auto-triggering, fewer ineffective efforts, and stable VT and ventilation.
Double triggering and breath stacking
- Pattern
- Two breaths occur with little or no expiration between them, increasing total VT.
- Likely causes
- Prolonged neural inspiration, high effort, inadequate flow, premature cycling, or reverse triggering.
- What to check
- Assess stacked VT, inspiratory duration, flow demand, effort, pain, acidosis, and sedation.
- Possible adjustment
- Inspiratory flow, pressure rise, inspiratory time or cycling criterion, and support level.
- After adjustment
- Separated breaths, acceptable VT and pressures, lower effort, and preserved gas exchange.
Insufficient inspiratory flow
- Pattern
- During volume control, the pressure curve becomes concave and inspiratory effort persists.
- Likely causes
- Insufficient flow, pressure rise, or support; pain, fever, anxiety, acidosis, or high ventilatory demand.
- What to check
- Compare delivered flow with patient demand and identify clinical causes of increased respiratory drive.
- Possible adjustment
- Flow pattern and peak flow, pressure rise, inspiratory time, support, and treatment of the cause.
- After adjustment
- Lower effort and distress, smoother pressure contour, acceptable VT and pressures, and adequate expiratory time.
Premature cycling
- Pattern
- Ventilator inspiration ends before patient effort, often causing immediate retriggering.
- Likely causes
- High flow-cycle threshold, short inspiratory time, or insufficient flow or support.
- What to check
- Compare the end of ventilator inspiration with patient effort; check leak, cycling criterion, inspiratory time, and VT.
- Possible adjustment
- Cycling criterion, inspiratory time, inspiratory flow, and support level.
- After adjustment
- Better breath matching without excessive VT, double triggering, or shortened expiration.
Delayed cycling
- Pattern
- Inspiratory flow continues after patient effort ends, with active expiration or a late pressure rise.
- Likely causes
- Low flow-cycle threshold, excessive support, leak, long inspiratory time, or obstruction.
- What to check
- Assess expiratory effort, leak, flow decay, respiratory mechanics, and support level.
- Possible adjustment
- Cycling criterion, inspiratory time, leak correction, pressure support, and pressure-rise setting.
- After adjustment
- Timely expiration, less expiratory effort, and adequate VT and ventilation.
Incomplete expiration
- Pattern
- Expiratory flow does not return to zero before the next inspiration.
- Likely causes
- Obstruction, long time constant, high minute ventilation, short expiratory time, or active expiration.
- What to check
- Assess resistance, ventilatory demand, expiratory time, haemodynamics, and total PEEP.
- Possible adjustment
- Increase expiratory time; adjust rate, inspiratory flow or time, support, and treatment of obstruction.
- After adjustment
- Improved lung emptying, lower total PEEP, better triggering, gas exchange, and haemodynamics.
Changes in Ppeak and Pplat
- Pattern
- Ppeak rises with stable Pplat, or both pressures rise together.
- Likely causes
- An isolated Ppeak rise suggests increased resistance; a Pplat rise suggests reduced compliance or increased total PEEP.
- What to check
- Repeat passive measurements and inspect the tube, circuit, airways, chest wall, lung volume, and patient position.
- Possible adjustment
- Treat airway or circuit obstruction; adjust VT, PEEP, position, and treatment of the underlying disease.
- After adjustment
- Repeat Ppeak, Pplat, total PEEP, Crs, and Rinsp and assess ventilation and haemodynamics.
Troubleshooting
Assess the patient, airway, circuit, and ventilator in parallel. If ventilation is ineffective or the patient is unstable, call for help and provide an immediately available alternative means of ventilation under the local airway-emergency pathway while the cause is sought.
- Sudden instability or absent ventilation
Confirm chest movement, oxygenation, airway position and patency, circuit continuity, gas supply, and exhalation. Separate disconnection or leak from tube/circuit obstruction, pneumothorax, severe bronchospasm, and loss of respiratory activity.
- High airway pressure
Use valid Ppeak–Pplat–PEEPtotal partitioning when the patient can be passive. Treat tube, secretion, bronchospasm, or circuit resistance when the rise is resistive; investigate lung/chest-wall volume, pleural or abdominal causes, position, and total PEEP when the elastic component rises.
- Low or changing expired volume
Check leak and disconnection first, then pressure limits, mechanics, inspiratory time, effort, cycling, and the active targeting scheme. In pressure control, falling VT may reveal worsening mechanics; in adaptive modes, changing pressure may be the controller response rather than the cause.
- Worsening oxygenation
Verify oxygen delivery and airway/circuit integrity, then assess recruitment loss, secretion or atelectasis, oedema, pneumothorax, haemodynamics, position, and ventilation–perfusion causes. Reassess the full oxygenation strategy and patient response, not the mode label alone.
- Apnea or loss of patient triggering
Immediately restore effective ventilation and evaluate central respiratory activity, sedation, neurological change, fatigue, over-assistance, intrinsic PEEP, trigger signal, and circuit continuity. Confirm that the active mandatory backup and alarms match the patient’s current risk.
Sources
1. Primary source
ISO 19223:2019 Lung ventilators and related equipment — Vocabulary and semantics
International Organization for Standardization · Edition 1; confirmed in 2024
Accessed Aug 1, 2026
2. Primary source
Invasive Ventilation and Use of Extracorporeal Procedures in Acute Respiratory Insufficiency — S3 Guideline
Association of the Scientific Medical Societies in Germany · Version 2.0, dated 2025-08-18
Accessed Aug 1, 2026
3. Primary source
ESICM guidelines on acute respiratory distress syndrome: definition, phenotyping and respiratory support strategies
European Society of Intensive Care Medicine · 2023
Accessed Aug 1, 2026
4. Primary source
An Official American Thoracic Society/American College of Chest Physicians Clinical Practice Guideline: Liberation from Mechanical Ventilation in Critically Ill Adults
American Thoracic Society and American College of Chest Physicians · 2017
Accessed Aug 1, 2026
5. Supporting source
A taxonomy for mechanical ventilation: 10 fundamental maxims
Respiratory Care · 2014
Accessed Aug 1, 2026
6. Supporting source
Respiratory Mechanics
Anesthesiology Clinics · 39(3):415–440
Accessed Aug 1, 2026
7. Supporting source
The basics of respiratory mechanics: ventilator-derived parameters
Annals of Translational Medicine · 6(19):376
Accessed Aug 1, 2026
8. Supporting source
Patient–Ventilator Dyssynchrony in Critically Ill Patients
Journal of Clinical Medicine · 10(19):4550; published 2021-09-30
Accessed Aug 1, 2026
9. Supporting source
HAMILTON-C6 Operator’s Manual, software version 2.0.x
Hamilton Medical · PN 624945.05; uploaded 2025-12-04
Accessed Aug 1, 2026
10. Supporting source
HAMILTON-C6 Operator’s Manual Addendum, software version 2.0.2
Hamilton Medical · PN 10203811/00; dated 2026-03-31
Accessed Aug 1, 2026
11. Supporting source
HAMILTON-C6 Technical specification for software version 2.0.x
Hamilton Medical · PN 689596.05
Accessed Aug 1, 2026
12. Supporting source
HAMILTON-C6 INTELLiVENT-ASV Operator’s Manual, software version 2.0.x
Hamilton Medical · PN 10172294/00; dated 2024-09-30
Accessed Aug 1, 2026
13. Supporting source
Evita V800 / V600 Instructions for Use, software 2.n
Dräger · PN 9511608; Edition 1, 2021-12 and higher
Accessed Aug 1, 2026
14. Supporting source
Evita V800 / V600 software 2.n options supplement
Dräger · PN 9513302; Edition 1, 2021-12 and higher
Accessed Aug 1, 2026
15. Supporting source
Make the most of your Servo-u
Getinge · Servo-u software 4.6; MX-8300 Rev04
Accessed Aug 1, 2026
16. Supporting source
Product Manuals — Puritan Bennett 840 PAV+ option catalogue
Medtronic · Page ©2026; PB840 PAV+ option addendum entries; manufacturer naming identity only
Accessed Aug 1, 2026
